- Info
- Specifications
- Certificates 1
Specifications
| Length of guide pipe components, m probe | 0.75 / 1.5 / 3 (measuring probe); 0.25 / 0.5 / 1 (connector) | ||
|---|---|---|---|
| Number of probes | from 1 to 10 | ||
| Guide pipe length, mm | from 750 to 30000 | ||
| Acceptable main inaccuracy of level measurement, mm | ±5 | ||
| Level measurement discreteness, mm: | 5 | ||
| Number of temperature measurement points, pcs | from 1 to 10 (determined by number of probes) | ||
| Temperature measurement accuracy, ℃ | ±0.5 (in range from -40 to +105); ±2 (in range from -50 to -40) | ||
| Monitored media features | temperature, ℃ | from -50 to +80 (explosive medium); from -50 to +100 (non-explosive medium) | |
| pressure, MPa, not more than | 1.6 / 2.5 (determined in accordance with types of float) | ||
| Number of control values of medium parameters (level, temperature, density, volume, weight, etc.) | 8 | ||
| Климатическое исполнение по ГОСТ 15150 | — | ||
| Ambient temperature range, ℃ | from -50 to +60 | ||
| Water and dust ingress protection degree as per IEC 60529 | IP66 (for housing); IP68 (for probes, connectors) | ||
| Explosion protection marking | 0/1Ex ia/db IIB T5 Ga/Gb X | ||
| Length of power and signal line, m, not more than length of signal line | 1500 | ||
| Output signal | SENS (SENS protocol), RS-485 (Modbus RTU protocol) | ||
| Power supply voltage (Up), V | from 6 to 50 | ||
| Power consumption, mW, not more than | 750 | ||
| Mean time between failures, hours, not less | 100000 | ||
| Weight, kg, not more than | 50 | ||
| Service life, years | 15 | ||